Free lab calculator
Standard curve calculator for linear and 4PL fits
Paste your standards and sample signals, choose a linear, through-the-origin or four-parameter logistic (4PL) model, and this standard curve calculator gives you the equation, R², a residual check and the concentration of every sample, multiplied by its dilution factor. Samples outside the standards are flagged, and the plot downloads as PNG or SVG.
- Free to calculate, no sign-up
- Runs in your browser
- Linear and 4PL models
- Flags samples outside the standards
- The residuals change sign only 2 times across 9 standards, which suggests the standards bend away from a straight line. Look at the residual plot, and consider the 4PL model or dropping standards beyond the linear range.
- 1 sample lies above the top standard. Dilute it further and measure again, rather than reporting an extrapolated value.
| Sample | Signal | From curve (µg/mL) | Dilution | Concentration (µg/mL) | Status |
|---|---|---|---|---|---|
| Lysate A | 0.612 | 554.2 | 5 | 2771 | Within the standards |
| Lysate B | 0.845 | 826.8 | 5 | 4134 | Within the standards |
| Lysate C | 1.164 | 1200 | 2 | 2400 | Within the standards |
| Lysate D | 2.218 | 2433 | 1 | 2433 | Extrapolated: the signal is above every standard. |
| Conc. (µg/mL) | Mean | SD | n | Fitted | Residual | Back-calc. | Recovery |
|---|---|---|---|---|---|---|---|
| 0 | 0.0925 | 0.002121 | 2 | 0.1383 | −0.0458 | −53.6 | n/a |
| 25 | 0.125 | 0 | 2 | 0.1597 | −0.0347 | −15.58 | −62.3% |
| 125 | 0.2265 | 0.002121 | 2 | 0.2452 | −0.0187 | 103.2 | 82.5% |
| 250 | 0.358 | 0.002828 | 2 | 0.352 | 0.006 | 257 | 103% |
| 500 | 0.6015 | 0.01061 | 2 | 0.5657 | 0.0358 | 541.9 | 108% |
| 750 | 0.8415 | 0.009192 | 2 | 0.7794 | 0.0621 | 822.7 | 110% |
| 1000 | 1.043 | 0.02192 | 2 | 0.9931 | 0.0504 | 1059 | 106% |
| 1500 | 1.426 | 0.001414 | 2 | 1.42 | 0.00557 | 1507 | 100% |
| 2000 | 1.787 | 0.01697 | 2 | 1.848 | −0.0608 | 1929 | 96.4% |
Recovery is the back-calculated concentration as a percentage of the nominal one; rows outside 80 to 120% are shaded. Check your assay's own acceptance limits.
Everything runs in your browser: your data is never uploaded.
How to use it
Calculate concentrations from a standard curve in three steps
Paste the standards, add your samples and pick a model, then read the curve, the fit and every concentration. Everything runs in your browser.
1Paste your standards
Copy the concentration column and the replicate signal columns from Excel and paste them in. A header row is fine; the box counts the standards and replicates it found.
2Add samples and choose a model
Enter each sample’s name, signal and dilution factor, then pick Linear, Through 0 or 4PL. The log axis spreads out standards that span several decades.
3Read the curve and the results
R² and the fitted parameters sit above the chart, and each sample is read off the curve. The tables below list every concentration; download PNG, SVG or CSV.
Five runs on the BCA and ELISA examples
Real results from the calculator: a straight line and a 4PL on the same BCA standards, the ELISA samples, and an S-shaped curve forced onto a line. Each caption gives the numbers the tool showed.





Guide
Making, checking and reading a standard curve
How a standard curve turns a signal into a concentration, which model to choose, what the 4PL parameters mean, how to check the fit, and how to handle dilutions and samples outside the curve.
How to use a standard curve to find a concentration
A standard curve links a measured signal, such as absorbance, fluorescence, luminescence or optical density, to known concentrations. You measure standards of known concentration in the same run as your samples, fit a line or curve through the standards, and read each sample’s concentration off that curve from its signal.
- Standards
- Read off the curve
- Extrapolated: dilute and re-run
- Run the standards on the same plate or in the same batch as the samples, with the same reagents and timing.
- Measure each standard in duplicate or triplicate. The calculator averages the replicates and shows their SD.
- Fit the curve with the signal on the y axis and concentration on the x axis, then invert it. For a straight line y = a + bx, the concentration is x = (y − a) / b.
- Multiply that result by the sample’s dilution factor to get the concentration in the original sample.
Bracket your samples
Choose standards whose signals span the signals you expect from your samples. A result is only as good as the part of the curve it is read from.
Linear, through the origin, or 4PL?
Pick the simplest model that follows your standards without leaving a pattern in the residuals.
Linear
y = a + bx
Through 0
y = bx, blank reads 0
4PL
S-shaped, e.g. ELISA
- Linear (y = a + bx): for assays whose signal rises in a straight line across the standards. The intercept a absorbs the signal of the blank.
- Through the origin (y = bx): only when a zero standard really reads zero, usually after you subtract the blank. Forcing the line through zero while the blank reads above it biases the low samples.
- Four-parameter logistic (4PL): for S-shaped curves that flatten at low and high concentrations, as immunoassays such as ELISA usually do across their working range. It needs standards at five or more concentrations.
- Colorimetric assays can bend at the top of their range. If a straight line leaves the residuals in an arch, drop the top standards or switch to the 4PL rather than trust a high R².
The four parameters of a 4PL standard curve
The 4PL here is y = D + (A − D) / (1 + (x / C)^B), an equivalent way of writing the model GraphPad Prism lists as “Sigmoidal, 4PL, X is concentration” for interpolating standard curves.
- A is the response at zero concentration, where the curve starts.
- D is the response the curve levels off at, at very high concentrations.
- C is the inflection point: the concentration that gives a response halfway between A and D, which GraphPad calls the EC50.
- B is the slope factor. Larger values give a steeper middle section. Here B stays positive, so a falling curve, as in a competitive ELISA, simply has A above D.
- The curve only gives concentrations for signals between A and D. A signal beyond either plateau has no answer, and the calculator says so instead of printing a number.
Checking the fit: R², residuals and back-calculated standards
R² says how much of the variation in the signal the curve explains, but a high R² can hide a curve of the wrong shape, especially when the highest standards dominate. Look at two more checks before you trust the results.
4PL: random scatter, the model fits
Straight line: an arch, the model has the wrong shape
- Residuals are each standard’s mean signal minus the fitted signal. They should scatter above and below zero; an arch or a trend means the model has the wrong shape.
- Back-calculated concentrations read each standard off the curve as if it were a sample. Recovery is that value as a percentage of the nominal concentration, and misfits usually show up first at the low end.
- R² is computed around the mean signal for all three models, so a line forced through zero can score lower than some spreadsheet trendlines report.
- Your kit insert or lab procedure sets the acceptance limits for R² and recovery. The calculator shades recoveries outside 80 to 120% as a prompt to look, not as a rule.
Dilution factors, replicates and samples outside the curve
The curve gives the concentration in the well or cuvette. The dilution factor takes you back to the original sample.
- Enter the total fold dilution: 10 for a 1:10 dilution, or 50 for a 1:5 dilution followed by a 1:10 dilution.
- Average replicate wells of a sample before you enter its signal, or enter each well on its own line and average the concentrations afterwards.
- A signal below the lowest standard or above the highest is extrapolated. The calculator still shows the number but flags it. Dilute a high sample further and measure it again instead of reporting the extrapolated value.
- With the 4PL model, a signal beyond a plateau gives no value at all. GraphPad Prism also leaves such results blank, because no concentration on the fitted curve gives that signal.
Blank subtraction
If you subtract the blank, subtract it from the standards and the samples alike. A blank-corrected zero standard then reads close to zero, which is when a line through the origin makes sense.
How this calculator fits the curve
All the arithmetic runs in your browser, and nothing you paste is uploaded.
- Replicate signals are averaged, and the curve is fitted to the mean of each standard.
- Straight lines use ordinary least squares. The 4PL uses the Levenberg-Marquardt method for nonlinear least squares, started from several initial guesses, keeping the best fit.
- On a log concentration axis a zero standard stays in the fit, although a log axis cannot show it.
- Rows whose first cell is not a number, such as a header, are skipped and listed under the box, and rows with the same concentration are pooled as replicates.
Frequently asked questions
Quick answers on calculating concentrations, R², lines through zero, 4PL curves, extrapolation and entering replicates.
How do you calculate concentration from a standard curve?
Fit a line through the standards, y = a + bx, then rearrange it to x = (y − a) / b and multiply by the dilution factor. With y = 0.00085x + 0.14 and a sample reading 0.61 that was diluted 5-fold, x = (0.61 − 0.14) / 0.00085 ≈ 553 in the well, or about 2,765 in the original sample, in the units of your standards.
What is a good R² value for a standard curve?
There is no single cutoff for every assay; your kit insert or lab procedure is the reference. R² alone can hide a curve of the wrong shape, so also check that the residuals scatter around zero and that the standards back-calculate close to their nominal values.
Should a standard curve go through zero?
Only if a zero standard really gives zero signal, for example after you subtract the blank from every well. Otherwise fit a line with an intercept, which absorbs the blank signal. Forcing the line through the origin while the blank reads above zero makes low samples read too high.
When should I use a 4PL instead of a linear standard curve?
Use a 4PL when the standards follow an S shape that flattens at the low or high end, as ELISA standards usually do. Use a straight line when the standards rise linearly across the range you need. If a straight line leaves an arch in the residuals, the data are telling you to switch.
Can I report a sample that falls outside the standard curve?
Not as a measured value. A signal above the top standard is extrapolated, so dilute the sample further and measure it again. A signal below the lowest standard is usually reported as below the range of the assay. The calculator flags both.
Why does a sample show “No value” with the 4PL model?
Its signal lies beyond one of the curve’s plateaus, A or D, and no concentration on the fitted curve gives that signal. It is usually a sample above the top of the assay range, which needs more dilution, or a reading below the blank.
How do I enter replicates and dilution factors?
For standards, paste the concentration in the first column and each replicate signal in the columns beside it, or one row per well; rows with the same concentration are pooled. For samples, enter a name, the signal and the fold dilution on each line, such as “Lysate A, 0.612, 5”.
Related tools
More free tools and chart generators for assay data.
IC50 calculator
Fit a four-parameter logistic curve to dose-response data and get the IC50 with an approximate 95% CI.
Regression plot generator
Turn a CSV or Excel file into a regression chart with the fitted line and its equation.
96 well plate template
Lay out standards, blanks and samples on a plate map before you pipette, and export it.
Error bar chart generator
Plot means with standard deviation, standard error or confidence intervals from your data.
Michaelis-Menten plot generator
Fit Km and Vmax to enzyme rate data and draw the saturation curve.
Plot digitizer
Pull data points out of a published standard curve or chart image, then export CSV.
Need the curve as a figure for a paper?
SciFigure chart generators turn CSV or Excel data into publication-style regression and dose-response charts.
Browse chart generators